COBAS E 411 IMMUNOASSAY ANALYZER E411 RACK 04775201001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-02-15 for COBAS E 411 IMMUNOASSAY ANALYZER E411 RACK 04775201001 manufactured by Roche Diagnostics.

Event Text Entries

[100451245] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[100451246] The customer experienced an issue over several days where the elecsys free psa immunoassay would be ordered in the cobas e 411 immunoassay analyzer software for patient samples and qc material where the test had not been requested and the result for free psa was identical for all patient samples. The field service representative could not find a cause. Operational and mechanical checks were acceptable. The customer powered down and rebooted the analyzer multiple times. The customer's laboratory information system (lis) was also restarted. The instrument was then running as expected and the problem seemed to be resolved. On (b)(6) 2018, the customer repeated all of the patient samples. Two examples were provided. Patient 1 original result was 39. 98 ng/ml and the repeat result was 1. 87 ng/ml. Patient 2 original result was 39. 98 ng/ml and the repeat result was 0. 488 ng/ml. No erroneous results were reported outside of the laboratory. The repeat results were believed to be correct. No patients were adversely affected. The free psa reagent lot number was 243142. The expiration date was requested but was not provided. The investigation confirmed the software issue which has been corrected in a new software version.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-00453
MDR Report Key7273535
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-02-15
Date of Report2018-02-15
Date of Event2018-01-26
Date Mfgr Received2018-01-30
Date Added to Maude2018-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS E 411 IMMUNOASSAY ANALYZER
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeLTJ
Date Received2018-02-15
Model NumberE411 RACK
Catalog Number04775201001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-15