OLYMPIC COOL-CAP 401701-501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-02-16 for OLYMPIC COOL-CAP 401701-501 manufactured by Natus Medical Inc..

Event Text Entries

[100510959] The cool cap is currently in the process of returning to the asc for further evaluation.
Patient Sequence No: 1, Text Type: N, H10


[100510960] The event involves a cool-cap unit with an e90 filling error. There has been no report of a death or serious injury that required treatment, a delay in treatment, and an environmental or safety concerns. A representative from an authorized service center (asc) contacted natus technical service concerning a cool cap with a filling error (e90). The customer reported that the pump had some faults and that the internal filter was cleaned prior to testing, but the filling error message was still present.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00102
MDR Report Key7277192
Report SourceUSER FACILITY
Date Received2018-02-16
Date of Report2018-01-18
Date Mfgr Received2018-01-18
Date Added to Maude2018-02-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANA SZUCS
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685133
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-02-16
Model Number401701-501
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC.
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-02-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.