CARDINAL HEALTH PERSONAL CLEANSING CLOTH, NON-FLUSHABLE, SCE 552AWS64

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-20 for CARDINAL HEALTH PERSONAL CLEANSING CLOTH, NON-FLUSHABLE, SCE 552AWS64 manufactured by Yangzhou Perfect Daily Chemicals.

Event Text Entries

[100773614] The device history record could not be reviewed as a lot number was not provided. Yangzhou perfect daily chemicals reviewed all their incoming, in-process and final inspection records for their last three lot numbers manufactured and all met specification. They reviewed their skin irritation testing report and skin sensitization testing report and both met specification. As a sample was not available for further investigation, the root cause could not be determined. There is no additional action taken at this time, but yangzhou perfect daily chemicals will continue to monitor the trend of this type of incident.
Patient Sequence No: 1, Text Type: N, H10


[100773615] The end user patient reported that wipes are causing skin irritation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423537-2018-00159
MDR Report Key7282578
Date Received2018-02-20
Date of Report2018-02-20
Date of Event2018-01-26
Date Facility Aware2018-01-26
Report Date2018-02-03
Date Reported to Mfgr2018-02-03
Date Mfgr Received2018-02-03
Date Added to Maude2018-02-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPATRICIA TUCKER
Manufacturer Street1500 WAUKEGAN RD
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer G1YANGZHOU PERFECT DAILY CHEMICALS
Manufacturer StreetHANGJI INDUSTRIAL PARK YANGZHOU
Manufacturer CityJIANGSU 225111
Manufacturer CountryCH
Manufacturer Postal Code225111
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCARDINAL HEALTH PERSONAL CLEANSING CLOTH, NON-FLUSHABLE, SCE
Generic NamePATIENT PERSONAL HYGIENE KIT
Product CodeNSB
Date Received2018-02-20
Model Number552AWS64
Catalog Number552AWS64
Lot NumberN/A
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerYANGZHOU PERFECT DAILY CHEMICALS
Manufacturer AddressHANGJI INDUSTRIAL PARK YANGZHOU JIANGSU 225111 CH 225111


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-02-20