TUR IRRIGATION SET 2C4005

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-02-20 for TUR IRRIGATION SET 2C4005 manufactured by Baxter Healthcare - Cartago.

Event Text Entries

[100526420] (b)(4). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[100526421] It was reported that the sterile packaging of three (3) y-type tur irrigation sets was open. This event was discovered before patient use during the kitting process. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[104097634] Two devices were received for evaluation. During visual examination, the pouch was observed open; therefore, the peel and seal width test was performed to evaluate the quality of the seal. The results were satisfactory due to the sealed is uniformed across the sealed area. Also, it was observed that the packaging material is correctly aligned. The reported condition was verified; however, the cause of the condition could not be determined. There are controls in place to detect and prevent an opening packaging in the manufacturing. The third device was not returned for evaluation; therefore, a device analysis could not be completed. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2018-00912
MDR Report Key7284494
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-02-20
Date of Report2018-03-23
Date Mfgr Received2018-03-12
Device Manufacturer Date2017-07-19
Date Added to Maude2018-02-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CARTAGO
Manufacturer StreetSEE H10 SEE H10
Manufacturer CityCARTAGO 30106
Manufacturer CountryCS
Manufacturer Postal Code30106
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTUR IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2018-02-20
Returned To Mfg2018-02-16
Model NumberNA
Catalog Number2C4005
Lot NumberR17G18057
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CARTAGO
Manufacturer AddressCARTAGO


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.