CENTERPIECE PLATE FIXATION SYSTEM G9010000274

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-02-22 for CENTERPIECE PLATE FIXATION SYSTEM G9010000274 manufactured by Warsaw Orthopedics.

Event Text Entries

[100793995] This product is not marketed in us but a similar device with catalog # 853-465, 510k# k050082 and upn (b)(4) is approved for sale in us. Product was not returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the r eported event. X-ray review: post op x ray c3-6 laminoplasty show placement of centerpiece laminoplasty plate. There is back out of the c4 lamina screw opposite to the facet screw. This is predictable for laminoplasty given fusion is not expected to occur. The operation is probably not necessary. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[100793996] It was reported that patient was pre-operatively diagnosed with cerebral spinal meningitis and underwent laminoplasty at c3-6. Post-op, the x-ray report revealed the screw back out at c4 and c6 on (b)(6) 2018. No patient complications reported as the result of the event
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1030489-2018-00257
MDR Report Key7288152
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-02-22
Date of Report2018-02-22
Date of Event2018-01-30
Date Mfgr Received2018-01-30
Date Added to Maude2018-02-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACIE ZIEMBA
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1WARSAW ORTHOPEDICS
Manufacturer Street2500 SILVEUS CROSSING
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCENTERPIECE PLATE FIXATION SYSTEM
Generic NameORTHOSIS, SPINE, PLATE, LAMINOPLASTY, METAL
Product CodeNQW
Date Received2018-02-22
Model NumberNA
Catalog NumberG9010000274
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerWARSAW ORTHOPEDICS
Manufacturer Address2500 SILVEUS CROSSING WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-22

© 2026 FDA.report