CHROMID? C. DIFFICILE AGAR 43871

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-02-23 for CHROMID? C. DIFFICILE AGAR 43871 manufactured by Biomerieux Spain.

Event Text Entries

[101140007] A veterinary customer from (b)(6) reported no growth on chromid? C. Difficile (lot 1006045090) while conducting an experiment. The customer stated the experiment was repeated using the customer's media (bhiss with added c. Diff supplement) and the chromid plates. Results were obtained with the customer's media but not with the chromid. The samples being tested were fecal specimens from animals that were infected with the c. Difficile 630 strain. The customer reported that samples with the c. Difficile 630 strain always grew well on previous lots of chromid so a reference strain was not used. A biom? Rieux internal investigation will be initiated. Although this event involved animal testing, the product chromid c. Difficile , can be used in a clinical setting and the situation of no growth can recur while testing a patient isolate.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00020
MDR Report Key7291954
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-02-23
Date of Report2018-04-19
Date Mfgr Received2018-03-26
Device Manufacturer Date2017-10-23
Date Added to Maude2018-02-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX SPAIN
Manufacturer StreetISAAC NEWTON, 6
Manufacturer CityTRES CANTOS, MADRID 28760
Manufacturer CountrySP
Manufacturer Postal Code28760
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? C. DIFFICILE AGAR
Generic NameCHROMID? C. DIFFICILE AGAR
Product CodeJSI
Date Received2018-02-23
Catalog Number43871
Lot Number1006045090
Device Expiration Date2018-02-05
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SPAIN
Manufacturer AddressISAAC NEWTON, 6 TRES CANTOS, MADRID 28760 SP 28760


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-23