ABBOTT VANCOMYCIN ASSAY 6E44-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-25 for ABBOTT VANCOMYCIN ASSAY 6E44-21 manufactured by Microgenics Corporation.

Event Text Entries

[101151144] The customer began using a new lot of vancomycin reagent lot 71368fd01. Prior to accepting the use of the new reagent lot, the customer calibrated and ran qc along with some archived patent samples to check the performance. Qc results were all in range with both lots. The customer had a couple of leftover patient samples with original values (on previous reagent lot 71346fd01) that were in the 31-34 ug/ml range. The customer decided to rerun leftover customer samples in parallel on the new and the old reagent lots. Customer observed low result of <1. 1ug/ml only on the new reagent lot. Customer diluted the sample and tested it again on the new reagent lot and obtained a result of 29 ug/ml customer concluded that the discrepancy is within the patient sample.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010939897-2018-00001
MDR Report Key7295374
Date Received2018-02-25
Date of Report2017-02-02
Date of Event2017-12-12
Date Mfgr Received2017-12-12
Device Manufacturer Date2017-08-07
Date Added to Maude2018-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationSERVICE AND TESTING PERSONNEL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DAVID SCHULTENOVER
Manufacturer Street46500 KATO ROAD
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal94538
Manufacturer Phone5109795000
Manufacturer G1MICROGENICS CORPORATION
Manufacturer Street46500 KATO ROAD
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal Code94538
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameABBOTT VANCOMYCIN ASSAY
Generic NameVANCOMYCIN TEST SYSTEM
Product CodeLEH
Date Received2018-02-25
Returned To Mfg2018-01-11
Model Number6E44-21
Catalog Number6E44-21
Lot Number71368FD01
Device Expiration Date2019-02-07
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROGENICS CORPORATION
Manufacturer Address46500 KATO ROAD FREMONT CA 94538 US 94538