CLINITRON RITEHITE C-8 ASSY P0800010019

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-26 for CLINITRON RITEHITE C-8 ASSY P0800010019 manufactured by Hill-rom Batesville.

Event Text Entries

[101153867] The hill-rom technician found the casters needed to be replaced. Per the hill-rom user manual, warning: patients may use the bed for support while entering or exiting; if the unit moves unexpectedly, patient injury could occur. When the unit is unattended, ensure that both brakes are locked. The brakes for the clinitron bed are located at the right, head end and the left, foot end of the unit. To apply the brakes, step on the lower end of the brake lever to lock the wheels. To release the brakes, apply inward pressure to the upper end of the brake lever. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2018. The technician replaced the casters to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[101153868] Hill-rom received a report from the account stating the brakes were not holding. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint # (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2018-00095
MDR Report Key7296709
Date Received2018-02-26
Date of Report2018-02-09
Date of Event2018-02-09
Date Mfgr Received2018-02-09
Date Added to Maude2018-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCLINITRON RITEHITE C-8 ASSY
Generic NameBED, AIR FLUIDIZED
Product CodeINX
Date Received2018-02-26
Model NumberP0800010019
OperatorOTHER CAREGIVERS
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.