75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-02-26 for 75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN manufactured by Charter Medical, Ltd..

Event Text Entries

[101245728] The bag was not returned for investigation. The user provided a picture of a bag, which indicated a film fracture below the bottom of one of the port tubes. A film crack the size of the one observed in the picture would have likely resulted in a detectable leak during filling of the bag. Therefore, it can be concluded that the crack likely occurred during handling, freezing, or thawing. The root cause of the film crack could not be conclusively determined. The film crack under the port tube suggests the bag may have encountered physical stress at some point, potentially while being handled when the bag was in the frozen state, although the exact point in the process where the film crack occurred is unknown. There are other causes that could potentially result in a film crack or tear: overfilling the bag. An overfilled bag could create resistance during insertion and removal of the bag from the metal cassette. Excessive air left inside the bag after filling, resulting in rapid expansion during thawing. Moisture on the exterior surface of the bag or cassette interior when the bag is inserted into the metal cassette. Moisture may cause the bag film to freeze to the interior cassette surface, thus potentially resulting in damage when the frozen bag is removed from the cassette. An unknown material anomaly. The product instruction for use sheet is provided with each cml-75ln device. Review of the ifu version provided with lot 150211 has the following precautions: "after freezing, do not handle excessively. Port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. " "do not overfill. " "remove as much air as possible from the container. " "ensure bag exterior and protective freezing cassettes are dry prior to initiating freezing protocol. Moisture on the exterior of the bag or on the cassette could cause adherence of the bag to the cassette resulting in difficulty of bag removal. "
Patient Sequence No: 1, Text Type: N, H10


[101245729] The user facility uses the 50-90 ml capacity cryobag cat #cml-75ln in their cell therapy program. They experienced a bag break involving a cml-75ln product from lot number #150211. When thawed, a leak was observed from a seam between the port and the body of the bag. The cells were salvaged and infused into a patient. There was no information available from the user regarding potential prophylactic treatments or any adverse event regarding this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00003
MDR Report Key7298035
Date Received2018-02-26
Date of Report2018-02-26
Date of Event2018-02-08
Date Mfgr Received2018-01-26
Device Manufacturer Date2017-08-03
Date Added to Maude2018-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic NameCELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2018-02-26
Model NumberCML-75LN
Catalog NumberCML-75LN
Lot Number150211
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103