MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-02-28 for PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE 04842464190 manufactured by Roche Diagnostics.
[101581990]
(b)(4). The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[101581991]
The customer complained of a false negative result for 1 patient tested for elecsys probnp ii immunoassay (probnp ii) in comparison to the creatinine result. The initial probnp ii result was < 5 pg/ml. The sample was repeated twice with results of < 5 pg/ml and < 5 pg/ml. The result of < 5 pg/ml was released outside of the laboratory, but was not believed to be correct since the patient had a creatinine result of 2. 5 mg/dl. There was no allegation of an adverse event. The customer used cobas 6000 e 601 module serial number (b)(4). The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2018-00611 |
MDR Report Key | 7302337 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-02-28 |
Date of Report | 2018-04-17 |
Date of Event | 2018-02-09 |
Date Mfgr Received | 2018-02-09 |
Date Added to Maude | 2018-02-28 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE |
Generic Name | TEST,NATRIURETIC PEPTIDE |
Product Code | NBC |
Date Received | 2018-02-28 |
Model Number | NA |
Catalog Number | 04842464190 |
Lot Number | 276345 |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-02-28 |