MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-02-28 for PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE 04842464190 manufactured by Roche Diagnostics.
[101581990]
(b)(4). The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[101581991]
The customer complained of a false negative result for 1 patient tested for elecsys probnp ii immunoassay (probnp ii) in comparison to the creatinine result. The initial probnp ii result was < 5 pg/ml. The sample was repeated twice with results of < 5 pg/ml and < 5 pg/ml. The result of < 5 pg/ml was released outside of the laboratory, but was not believed to be correct since the patient had a creatinine result of 2. 5 mg/dl. There was no allegation of an adverse event. The customer used cobas 6000 e 601 module serial number (b)(4). The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1823260-2018-00611 |
| MDR Report Key | 7302337 |
| Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
| Date Received | 2018-02-28 |
| Date of Report | 2018-04-17 |
| Date of Event | 2018-02-09 |
| Date Mfgr Received | 2018-02-09 |
| Date Added to Maude | 2018-02-28 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | NA MICHAEL LESLIE |
| Manufacturer Street | 9115 HAGUE ROAD NA |
| Manufacturer City | INDIANAPOLIS IN 46250 |
| Manufacturer Country | US |
| Manufacturer Postal | 46250 |
| Manufacturer Phone | 3175214343 |
| Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
| Manufacturer Street | SANDHOFERSTRASSE 116 NA |
| Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
| Manufacturer Country | GM |
| Manufacturer Postal Code | 68305 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Removal Correction Number | NA |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE |
| Generic Name | TEST,NATRIURETIC PEPTIDE |
| Product Code | NBC |
| Date Received | 2018-02-28 |
| Model Number | NA |
| Catalog Number | 04842464190 |
| Lot Number | 276345 |
| ID Number | NA |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Age | DA |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | ROCHE DIAGNOSTICS |
| Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2018-02-28 |