[101583486] The dymax adhesive was chipped off at two locations: one ( about 0. 14 mm) at distal end of the angle cover, about 7 mm from the distal end of the scope; the other one at the proximal end of angle cover, about 45 mm from distal end of the scope. There is an external cut in the angle cover that is 20 mm from the distal tip, causing an angle cover leak.
Patient Sequence No: 1, Text Type: N, H10


[101583487] Allegedly, video scope was used during a bronchoscopy for double lumen placement, the doctor noticed that a fragment or small piece of a blue material had come loose from the scope and was retained in patient's trachea. They brought in another scope to complete the procedure. The chief of anesthesia confirmed that physician does not have plans to remove the piece that stayed in the patient. The doctor believes that patient will cough up the piece.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1221826-2018-00016
MDR Report Key7305006
Date Received2018-02-28
Date of Report2018-02-01
Date of Event2018-02-01
Date Mfgr Received2018-02-01
Device Manufacturer Date2014-12-09
Date Added to Maude2018-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO CA 902455017
Manufacturer CountryUS
Manufacturer Postal902455017
Manufacturer Phone4242188100
Manufacturer G1KARL STORZ ENDOVISION
Manufacturer Street91 CARPENTER HILL ROAD
Manufacturer CityCHARLTON MA 01507
Manufacturer CountryUS
Manufacturer Postal Code01507
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFLEXIBLE INTUBATION VIDEO ENDOSCOPE
Generic NameVIDEO ENDOSCOPE
Product CodeCAL
Date Received2018-02-28
Returned To Mfg2018-02-20
Model Number11302BDX
Catalog Number11302BDX
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ ENDOVISION
Manufacturer Address91 CARPENTER HILL ROAD CHARLTON MA 01507 US 01507


Patients

Patient NumberTreatmentOutcomeDate
10 2018-02-28