TOOTHETTE? ORAL CARE DISPOSABLE ORAL SWAB 6071

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-03-01 for TOOTHETTE? ORAL CARE DISPOSABLE ORAL SWAB 6071 manufactured by Sage Products Llc.

Event Text Entries

[101634767] Reporter discarded the involved product and did not provide photographs or lot information. Due to the lack of lot information, a product history record could not be reviewed. Reporter stated that some green foam remained on the swab handle. This would indicate that the foam head was glued to the swab handle appropriately. Reporter stated the patient bit down on the green foam and caused it to disengage and no bite block was used. Per packaging instructions, it states, "use a bite block when performing oral care on patients with altered levels of consciousness, or those who cannot comprehend commands. " the root cause of the reported condition could be attributed to the patient's biting of the foam head.
Patient Sequence No: 1, Text Type: N, H10


[101634768] Report received of a foam disengagement due to biting. On (b)(6) 2018, oral care was performed on a male patient. The reporter stated the patient bit down on the green foam and the green foam disengaged from the stick. The reporter stated the patient swallowed the disengaged foam piece. The reporter stated no injury occurred to the patient or nurse who performed the oral care. The reporter did not know if the patient suffered from an altered level of consciousness or was unable to comprehend commands. The reporter stated the adverse event occurred during initial use of the product and a bite block was not in place. The reporter stated following foam disengagement, green foam pieces were evident on the oral swab stick. The device was discarded and no pictures were available. Although requested, no additional information provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419181-2018-00001
MDR Report Key7307110
Report SourceHEALTH PROFESSIONAL
Date Received2018-03-01
Date of Report2018-03-01
Date of Event2018-02-06
Date Mfgr Received2018-02-06
Date Added to Maude2018-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. HEATHER SIRIANNI
Manufacturer Street3909 THREE OAKS ROAD
Manufacturer CityCARY IL 60013
Manufacturer CountryUS
Manufacturer Postal60013
Manufacturer Phone8154554700
Manufacturer G1SAGE PRODUCTS LLC
Manufacturer Street3909 THREE OAKS ROAD
Manufacturer CityCARY IL 60013
Manufacturer CountryUS
Manufacturer Postal Code60013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOOTHETTE? ORAL CARE DISPOSABLE ORAL SWAB
Generic NameORAL SWAB
Product CodeKXF
Date Received2018-03-01
Model Number6071
Catalog Number6071
Lot NumberUNKNOWN
OperatorNURSE
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSAGE PRODUCTS LLC
Manufacturer Address3909 THREE OAKS ROAD CARY IL 60013 US 60013


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-01