ABL800 SERIES ABL825 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2006-06-19 for ABL800 SERIES ABL825 * manufactured by Radiometer Medical Aps.

Event Text Entries

[462289] After treatment of four patients based on measurements from the bloodgas analyser abl825, it was observed that all the four measurement results used for the treatment, were identical with the measurement conducted just before the measurements of the four patients. No harm to the patients has been reported. The problem is caused by a sw bug.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2006-00030
MDR Report Key730804
Report Source01
Date Received2006-06-19
Date of Report2006-06-16
Date of Event2006-05-16
Date Mfgr Received2006-05-17
Date Added to Maude2006-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactGERT NIELSEN
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ DK-2700
Manufacturer CountryDA
Manufacturer PostalDK-2700
Manufacturer Phone8273057
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABL800 SERIES
Generic Name*
Product CodeCCE
Date Received2006-06-19
Model NumberABL825
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key719092
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAKANDEVEJ 21 BRONSHOJ DA DK-2700
Baseline Brand NameABL700SERIES AND ABL800SERIES
Baseline Generic NameBLOOD GAZ ANALYZER
Baseline Model NoABL825
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2006-06-19

© 2026 FDA.report