CHROMID? STRB AGAR 43461

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-02 for CHROMID? STRB AGAR 43461 manufactured by Biomerieux Sa.

Event Text Entries

[101697817] A customer in (b)(6) notified biom? Rieux of media shrinkage that lead to a delayed result greater than 24 hours with chromid? Strb agar (reference 43461). The customer reported that the media shrinkage was observed on some plates after incubation. However, this issue has caused delayed results greater than 24 hours due to repeat testing. There is no indication or report from the laboratory or physician that the discrepant result led to any adverse event related to a patient's state of health. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00025
MDR Report Key7309693
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-02
Date of Report2018-06-04
Date Mfgr Received2018-05-07
Device Manufacturer Date2017-10-17
Date Added to Maude2018-03-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX SA
Manufacturer Street5 RUE DES AQUEDUCS 69290 CRAPONNE
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? STRB AGAR
Generic NameCHROMID? STRB AGAR
Product CodePQZ
Date Received2018-03-02
Catalog Number43461
Lot Number1006032880
Device Expiration Date2018-02-08
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address5 RUE DES AQUEDUCS 69290 CRAPONNE FR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-02

© 2026 FDA.report