STANDARD POLYSOMNOGRAPH WITH ELECTROENCEPHALOGRAPH SOMNOSTAR Z4 SYSTEM VER 10 23586-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-03-05 for STANDARD POLYSOMNOGRAPH WITH ELECTROENCEPHALOGRAPH SOMNOSTAR Z4 SYSTEM VER 10 23586-001 manufactured by Vyaire Medical, Inc.

Event Text Entries

[101511998] Vyaire identifier: (b)(4). The suspect component has been returned to vyaire and further evaluation is pending. Upon completion of the final evaluation, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[101511999] The customer reported while using the somnostar device, the monitor will not maintain a picture and it gave him (the user of the device) an electrical shock feeling. The user reported that there was no harm or injury, and no medical intervention was sought. The customer stated there was no patient involvement associated with this reported event.
Patient Sequence No: 1, Text Type: D, B5


[114592643] The vyaire medical failure analysis laboratory received the suspect component, a dell flat panel monitor, and evaluated the component. An evaluation of the component duplicated the reported issue and found that the part is nonfunctional, but there is no occurrence of electrical shock.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2021710-2018-07474
MDR Report Key7313311
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-03-05
Date of Report2018-03-20
Date of Event2018-02-07
Date Mfgr Received2018-03-08
Device Manufacturer Date2015-07-01
Date Added to Maude2018-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN GRAF
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer G1VYAIRE MEDICAL, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTANDARD POLYSOMNOGRAPH WITH ELECTROENCEPHALOGRAPH
Generic NameSOMNO STAR & SERIES SLEEP SYSTEM
Product CodeOLV
Date Received2018-03-05
Returned To Mfg2018-03-01
Model NumberSOMNOSTAR Z4 SYSTEM VER 10
Catalog Number23586-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-03-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.