MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05 report with the FDA on 2006-06-29 for FRESENIUS DIALYSIS DELIVERY SYSTEM 2008H NA manufactured by Fresenius Medical Care North America.
[483224]
A dialysis facility reported that a patient gained weight during a hemodialysis treatment. There were reportedly low tmp alarms that had to be reset every 5 minutes. About 3 hrs into the treatment, the patient c/o discomfort and tightness in her arms and fingers. The staff noticed that her face was swollen. Treatment was discontinued. The patient was weighed and her post dialysis weight indicated a weight gain of 5. 7 kg. She was taken to the hospital, dialyzed and discharged the next day.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2937457-2006-00004 |
MDR Report Key | 731685 |
Report Source | 00,05 |
Date Received | 2006-06-29 |
Date of Report | 2006-05-30 |
Date of Event | 2006-05-26 |
Date Facility Aware | 2006-05-26 |
Report Date | 2006-05-30 |
Date Reported to Mfgr | 2006-05-30 |
Date Mfgr Received | 2006-05-30 |
Device Manufacturer Date | 2001-01-01 |
Date Added to Maude | 2006-07-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | ELVY DIZON, BSN, RN, CNN |
Manufacturer Street | 2637 SHADELANDS DRIVE |
Manufacturer City | WALNUT CREEK CA 94598 |
Manufacturer Country | US |
Manufacturer Postal | 94598 |
Manufacturer Phone | 9252950200 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | FRESENIUS DIALYSIS DELIVERY SYSTEM |
Generic Name | HEMODIALYSIS MACHINE |
Product Code | FIL |
Date Received | 2006-06-29 |
Model Number | 2008H |
Catalog Number | NA |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | 5 YR |
Device Eval'ed by Mfgr | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 719973 |
Manufacturer | FRESENIUS MEDICAL CARE NORTH AMERICA |
Manufacturer Address | * WALNUT CREEK CA * US |
Baseline Brand Name | FRESENIUS DIALYSIS DELIVERY SYSTEM |
Baseline Generic Name | HEMODIALYSIS MACHINE |
Baseline Model No | 2008H |
Baseline Catalog No | F002-30 |
Baseline ID | NA |
Baseline Device Family | NA |
Baseline Shelf Life Contained | A |
Baseline PMA Flag | N |
Baseline 510K PMN | Y |
Premarket Notification | K921456 |
Baseline Preamendment | N |
Baseline Transitional | N |
510k Exempt | N |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Hospitalization | 2006-06-29 |