MCP SZ. 30 PROXIMAL WW MCP-100-30P-WW

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-03-06 for MCP SZ. 30 PROXIMAL WW MCP-100-30P-WW manufactured by Ascension Orthopedics.

Event Text Entries

[101660554] The device involved in the reported incident has been received for evaluation. An investigation has been initiated based on the reported information.
Patient Sequence No: 1, Text Type: N, H10


[101660555] It was reported that the product was implanted for 7 years and signs of excessive wear was noted during a revision surgery scheduled due to squeaking and poor functionality. No incidents during the revision surgery, the implant was replaced and the surgeon was satisfied with the result.
Patient Sequence No: 1, Text Type: D, B5


[131597123] The part was returned and underwent failure analysis. Both fracture and abrasive wear damage was evident on articulating surface of implant. Fracture damage suggest multiple events due to demanding hand function or trauma. Wear suggests subluxation or dislocation of prosthesis. However, it is not possible to identify cause of fracture and wear. Based on the information available to date, the root cause for the reported event cannot be determined. Possible causes include trauma. If additional information is received, the investigation will be updated.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1651501-2018-00015
MDR Report Key7318241
Report SourceHEALTH PROFESSIONAL
Date Received2018-03-06
Date of Report2017-12-18
Date Mfgr Received2018-03-19
Date Added to Maude2018-03-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER VIVIAN NELSON
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1ASCENSION ORTHOPEDICS
Manufacturer Street8700 CAMERON ROAD #100
Manufacturer CityAUSTIN TX 78754
Manufacturer CountryUS
Manufacturer Postal Code78754
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMCP SZ. 30 PROXIMAL WW
Generic NameMCP
Product CodeNEG
Date Received2018-03-06
Returned To Mfg2018-02-02
Catalog NumberMCP-100-30P-WW
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerASCENSION ORTHOPEDICS
Manufacturer Address8700 CAMERON ROAD #100 8700 CAMERON ROAD #100 AUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-03-06

© 2026 FDA.report