ACCUDENT XD TRAY MATERIAL 673471

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-03-07 for ACCUDENT XD TRAY MATERIAL 673471 manufactured by Ivoclar Vivadent, Inc..

Event Text Entries

[101703701] Accudent xd instruction for use contains the word "spearmint" under composition of tray material and under composition of syringe material. Due to the presence of spearmint in accudent xd, the dental office should not have used the product on a patient with a mint allergy. Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[101703702] Patient reported a mint allergy in her medical history. The doctor's staff checked packaging and labeling of accudent xd and could not find any evidence of the word "mint". As a result, the dental staff assumed the product was safe to use for this patient. A few hours after exposure, the patient had significant respiratory symptoms, and developed painful ulcers on her lips and buccal mucosa. The patient managed her reaction with benadryl. Accudent xd ifu does contain the word "spearmint" under composition.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1316092-2018-00002
MDR Report Key7319882
Report SourceHEALTH PROFESSIONAL
Date Received2018-03-07
Date of Report2018-03-07
Date of Event2018-02-20
Date Mfgr Received2018-02-27
Date Added to Maude2018-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. DONNA HARTNETT
Manufacturer Street175 PINEVIEW DRIVE
Manufacturer CityAMHERST NY 14228
Manufacturer CountryUS
Manufacturer Postal14228
Manufacturer Phone7166912260
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACCUDENT XD TRAY MATERIAL
Generic NameMATERIAL, IMPRESSION
Product CodeELW
Date Received2018-03-07
Catalog Number673471
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerIVOCLAR VIVADENT, INC.
Manufacturer Address175 PINEVIEW DRIVE AMHERST NY 14228 US 14228


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-03-07