[102221981] A call/complaint was received indicating the "collar broken under chin". The sample has not been received at this time to be able to complete a thorough investigation. The device history record was reviewed and no issues were found. Inventory was inspected and no issues were found. The root cause was unable to be determined, due to the absence of the sample the investigator could not confirm the reported issue. If a sample is received the complaint will be reopened and updated as necessary. Between, january 21, 2016 - march 2, 2018, there have been a total of 2 collars reported to be malfunctioning for a percentage of (b)(4) units sold, of this product 1044-40 cervical collar, emt select, one piece. No further information is available at this time. We will provide a follow up report if additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10


[102221982] Quality issue details: date of occurrence: (b)(6) 2018. When did quality issue occur? Before use. Who was using or operating the product when the quality issue occurred? Health professional. Was a medical procedure involved? No. Name of medical procedure: not applicable. Did the quality issue cause a delay in the medical procedure? Not applicable. Detailed description of quality issue: broken under chin. How was the quality issue was identified? By visual inspection. How was the product being used? Collar. Was it the initial use of the product? Yes. Was the product modified from the original condition supplied by deroyal? No. Was the product connected to or used in conjunction with other devices or equipment? No. Outcome details: outcome(s) attributed to quality issue: none. Person(s) affected by outcome(s) checked above: none. Known pre-existing condition(s) of person(s) affected: none specified. Was the incident reported to the fda? No. Detailed description of outcome(s), including information regarding injury or any additional treatment/intervention required: na.
Patient Sequence No: 1, Text Type: D, B5


[120879621] Engineering completed testing on the product and found during the backward bend test the adult collar failed at an average displacement of 7. 14mm, and the pediatric collar failed at an average displacement of 3. 83mm. Whereas in the forward bend test, both the adult and pediatric collars were able to reach 10mm of displacement without breaking. In conclusion, the bridge of the collars are most likely failing, when bent backwards, as was performed in the backward bend test. Based on the testing performed by the engineering department, the product will perform properly if used as intended and no excessive force is applied during application or use. Based on the testing completed and conclusion reached by all concerned departments, as a preventive action communication was sent outlining plans to move forward with a change to the resin in order to safeguard against possible misuse or force beyond the breaking point of the material during unusual application. This will serve to further improve the product. Between, january 1, 2016 - august 3, 2018 there have been a total of 12 collars reported to be malfunctioning and (b)(4) units sold for a percentage of (b)(4)%, of this product 14-103a-00 neck orthosis, semi-rigid. The investigation is complete at this time. We will provide a follow up report if additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006851902-2018-00002
MDR Report Key7323931
Report SourceUSER FACILITY
Date Received2018-03-08
Date of Report2018-02-12
Date of Event2018-02-12
Date Mfgr Received2018-02-12
Date Added to Maude2018-03-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MELISSA LOGSDON
Manufacturer Street200 DEBUSK LANE
Manufacturer CityPOWELL TN 37849
Manufacturer CountryUS
Manufacturer Postal37849
Manufacturer Phone8653626157
Manufacturer G1DEROYAL INTERCONTINENTAL , S.R.L
Manufacturer StreetKM 7, AUTOPISTA JOAQUIN BALAGUER, PISANO FREE ZONE, 18
Manufacturer CitySANTIAGO,
Manufacturer CountryDR
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDEROYAL
Generic NameCERVICAL COLLAR
Product CodeIQK
Date Received2018-03-08
Returned To Mfg2018-04-25
Model Number14-103A-00
Lot Number45374874
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEROYAL INTERCONTINENTAL , S.R.L
Manufacturer AddressKM 7, AUTOPISTA JOAQUIN BALAGUER, PISANO FREE ZONE, 18 SANTIAGO, DR