OPMI PICO ON S100 WALL MOUNT 000000-1403-544

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,user facility report with the FDA on 2018-03-08 for OPMI PICO ON S100 WALL MOUNT 000000-1403-544 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[101993274] The manufacture made attempts to access the microscope system for evaluation without success. Based on pictures provided by the hcp, the manufacturer determined the root cause of the malfunction to be incorrect screw or/and wall plug which was used to install the wall mount plate to the wall. The hcp informed the manufacturer that the microscope system was removed from its original installation site and was mounted at a different location. The reinstallation was performed by a 3rd party service provider.
Patient Sequence No: 1, Text Type: N, H10


[101993275] A healthcare professional (hcp) reported that the supporting wall mount plate, on which the opmi pico with s100 wall mount microscope system was mounted, came loose from the wall and the microscope system dropped on a bed below. The incident happened after the surgery had already been completed. No one was using the microscope at the time and no patient was involved. There was no injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2018-00003
MDR Report Key7325842
Report SourceFOREIGN,USER FACILITY
Date Received2018-03-08
Date of Report2018-02-07
Date of Event2018-02-07
Date Mfgr Received2018-02-07
Device Manufacturer Date2012-10-31
Date Added to Maude2018-03-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 WALL MOUNT
Generic NameMICROSCOPE, SURGICAL, GENERAL & PLASTIC SURGERY
Product CodeFSO
Date Received2018-03-08
Model NumberNA
Catalog Number000000-1403-544
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-03-08