MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2018-03-09 for TRIDENT ENDOSCOPIC ULTRASOUND BIOPSY NEEDLE(FNB) EUS-22-1-N manufactured by Micro-tech(nanjing) Co.,ltd..
[102092040]
Corrective action and preventive action: we will modify the design drawings to shorten the length of the developing area of the needle point engraving and avoid the force position of the pliers. -due date 03/29/2018. We will modify the inspection drawings, strengthen the toughness of the tube in the area of the needle point carving, and reduce the risk of the bad products entering the production -due date 04/16/2018.
Patient Sequence No: 1, Text Type: N, H10
[102092041]
Pre-market evaluation: clinician used competitive device that could not puncture tumor. They used micro-tech eus-fnb and were able to puncture. They went in a 2nd time with the same micro-tech needle to biopsy and when pulling the eus out, the needle had broken off and lodged into a side of the pancreas. The staff tried to remove the needle with forceps, but were unable to. The patient was admitted into a local hospital to have it removed.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004837686-2018-00001 |
MDR Report Key | 7327666 |
Report Source | COMPANY REPRESENTATIVE,DISTRI |
Date Received | 2018-03-09 |
Date of Report | 2018-03-06 |
Date of Event | 2018-02-13 |
Date Mfgr Received | 2018-02-14 |
Device Manufacturer Date | 2017-12-26 |
Date Added to Maude | 2018-03-09 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS BECKY LI |
Manufacturer Street | NO.10 GAOKE THIRD ROAD NANJING NATIONAL HI-TECH, INDU |
Manufacturer City | NANJING, JIANG SU 210032 |
Manufacturer Country | CH |
Manufacturer Postal | 210032 |
Manufacturer G1 | MICRO-TECH(NANJING) CO.,LTD. |
Manufacturer Street | NO.10 GAOKE THIRD ROAD NANJING NATIONAL HI-TECH, INDU |
Manufacturer City | NANJING, JIANG SU 210032 |
Manufacturer Country | CH |
Manufacturer Postal Code | 210032 |
Single Use | 3 |
Remedial Action | RC |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TRIDENT ENDOSCOPIC ULTRASOUND BIOPSY NEEDLE(FNB) |
Generic Name | ENDOSCOPIC ULTRASOUND NEEDLE |
Product Code | ODG |
Date Received | 2018-03-09 |
Returned To Mfg | 2018-02-23 |
Catalog Number | EUS-22-1-N |
Lot Number | M171226171 |
Device Expiration Date | 2019-12-25 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MICRO-TECH(NANJING) CO.,LTD. |
Manufacturer Address | NO.10 GAOKE THIRD ROAD NANJING NATIONAL HI-TECH, INDU NANJING, JIANG SU 210032 CH 210032 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Hospitalization | 2018-03-09 |