MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-03-09 for POSEY BED 8070 manufactured by Posey Products Llc.
[102215117]
Evaluation of the returned canopy found one element missing on the right window drainage port and one element missing on the molding zipper a patient access area. If one zipper element is missing or pulled apart on a patient accessible zipper it would allow the zipper to be initially closed. A patient can later open the zipper by separating the zipper at the location of the missing element. Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper. Posey beds are multi-use, serviceable items. As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use. If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing. Service issues are trended and reviewed by management on a monthly basis. As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted. No corrective or preventative actions are necessary at this time. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[102215118]
Customer reported the zipper teeth are missing from the right side panel of the canopy. The date the issue was discovered is unknown. No patient injury was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2020362-2018-00020 |
MDR Report Key | 7327795 |
Date Received | 2018-03-09 |
Date of Report | 2018-02-27 |
Date Mfgr Received | 2018-02-19 |
Device Manufacturer Date | 2014-05-02 |
Date Added to Maude | 2018-03-09 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | WILLIAM HINCY |
Manufacturer Street | POSEY PRODUCTS LLC 5635 PECK ROAD |
Manufacturer City | ARCADIA CA 91006 |
Manufacturer Country | US |
Manufacturer Postal | 91006 |
Manufacturer Phone | 6264433143 |
Manufacturer G1 | POSEY S DE RL DE CV |
Manufacturer Street | CAM. ANTIGUA A TECATE #16760 I COLONIA NI |
Manufacturer City | TIJUANA, BAJA CALIFORNIA 22120 |
Manufacturer Country | MX |
Manufacturer Postal Code | 22120 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 0 |
Brand Name | POSEY BED 8070 |
Generic Name | PATIENT BED WITH CANOPY/RESTRAINTS |
Product Code | OYS |
Date Received | 2018-03-09 |
Returned To Mfg | 2018-02-19 |
Model Number | 8070 |
Catalog Number | 8070 |
Lot Number | NA |
Operator | OTHER HEALTH CARE PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | POSEY PRODUCTS LLC |
Manufacturer Address | 5635 PECK ROAD ARCADIA CA 91006 US 91006 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-03-09 |