ELECSYS FREE PSA IMMUNOASSAY

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-12 for ELECSYS FREE PSA IMMUNOASSAY manufactured by Roche Diagnostics.

Event Text Entries

[102450164] (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[102450165] The customer complained of discrepant results for 1 patient sample tested for elecsys total psa immunoassay (total psa) and elecsys free psa immunoassay (free psa) on a cobas 6000 e 601 module. This medwatch will cover free psa. Refer to medwatch with (b)(6) for information on the total psa erroneous results. (b)(6). No incorrect results were reported outside of the laboratory. There was no allegation that an adverse event occurred. The e601 module serial number used for the initial results was (b)(4). The serial number for the e601 module used for repeat testing was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-00774
MDR Report Key7331221
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-12
Date of Report2018-05-22
Date of Event2018-02-19
Date Mfgr Received2018-02-22
Date Added to Maude2018-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS FREE PSA IMMUNOASSAY
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2018-03-12
Model NumberNA
Catalog NumberASKU
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.