TOTAL PSA, TOTAL (FREE + COMPLEXED) PSA - PROSTATE-SPECIFIC ANTIGEN (TPSA) 04641655190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-12 for TOTAL PSA, TOTAL (FREE + COMPLEXED) PSA - PROSTATE-SPECIFIC ANTIGEN (TPSA) 04641655190 manufactured by Roche Diagnostics.

Event Text Entries

[102336450] This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[102336451] The customer complained of discrepant results for 1 patient sample tested for elecsys total psa immunoassay (total psa) and elecsys free psa immunoassay (free psa) on a cobas 6000 e 601 module. This medwatch will cover total psa. Refer to medwatch with patient identifier (b)(6) for information on the free psa erroneous results. The initial total psa result from the e601 module was 0. 583 ng/ml. The sample was repeated on a different e601 module and the result was 0. 584 ng/ml. The initial free psa result from the e601 module was 1. 44 ng/ml. The sample was repeated on a different e601 module and the result was 1. 46 ng/ml. The same sample was repeated by an unspecified chemiluminescence method and the total psa result was 6. 71 ng/ml. No incorrect results were reported outside of the laboratory. There was no allegation that an adverse event occurred. The e601 module serial number used for the initial results was (b)(4). The serial number for the e601 module used for repeat testing was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-00773
MDR Report Key7331228
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-12
Date of Report2018-05-22
Date of Event2018-02-19
Date Mfgr Received2018-02-22
Date Added to Maude2018-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTOTAL PSA, TOTAL (FREE + COMPLEXED) PSA - PROSTATE-SPECIFIC ANTIGEN (TPSA)
Generic NameTOTAL,PROSTATE SPECIFIC ANTIGEN(NONCOMPLEXED&COMPLEXED) FOR DETECTION OF PROS
Product CodeMTF
Date Received2018-03-12
Model NumberNA
Catalog Number04641655190
Lot Number247045
ID NumberNA
Device Expiration Date2018-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.