TUR IRRIGATION SET 2C4041

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-03-12 for TUR IRRIGATION SET 2C4041 manufactured by Baxter Healthcare - Cartago.

Event Text Entries

[102124515] Contact office address: (b)(4). The device has been received and the evaluation is in progress. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[102124516] It was reported that a y-type tur/bladder irrigation set leaked after priming. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[105394020] The device was received for evaluation. Visual inspection did not identify any abnormalities that could have contributed to the reported condition. Pressure test and clear passage under water was performed with no issues noted. The reported condition was not verified. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2018-01288
MDR Report Key7331364
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-03-12
Date of Report2018-04-11
Date of Event2018-02-14
Date Mfgr Received2018-04-03
Date Added to Maude2018-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CARTAGO
Manufacturer StreetSEE H10 SEE H10
Manufacturer CityCARTAGO 30106
Manufacturer CountryCS
Manufacturer Postal Code30106
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTUR IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2018-03-12
Returned To Mfg2018-03-05
Model NumberNA
Catalog Number2C4041
Lot NumberASKU
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CARTAGO
Manufacturer AddressCARTAGO


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.