TOSOH HLC-723G8 ANALYZER G8 021560

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-03-12 for TOSOH HLC-723G8 ANALYZER G8 021560 manufactured by Tosoh Corporation.

Event Text Entries

[102446540] A field service engineer was dispatched (b)(6) 2016 to the address the reported event. The fse replaced the diaphram behind the pump plunger that was leaking from the wash chamber. There was no further action needed by field service. The g8 analyzer is in good working condition and was functioning as intended and the issue cleared. No further action required by field service. The most probable cause of the reported event the teflon diaphragm failed. (b)(4). This report is being submitted due to a retrospective review conducted under capa (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[102446541] On (b)(6) 2016 a customer reported a leak from the panel filter holder with the g8 analyzer. The customer is unable to run hba1c in the patient samples. A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting of hba1c patient results. There was no indication of any patient intervention or adverse health consequences due to the delay in reporting.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8031673-2018-01851
MDR Report Key7331394
Report SourceHEALTH PROFESSIONAL
Date Received2018-03-12
Date of Report2018-03-12
Date Facility Aware2016-05-04
Report Date2018-03-12
Date Reported to FDA2018-03-12
Date Reported to Mfgr2018-03-12
Date Mfgr Received2016-05-04
Device Manufacturer Date2011-03-01
Date Added to Maude2018-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS DORIA ESQUIVEL
Manufacturer Street6000 SHORELINE COURT SUITE 101
Manufacturer CitySOUTH SAN FRANCISCO CA 94080
Manufacturer CountryUS
Manufacturer Postal94080
Manufacturer Phone6506368123
Manufacturer G1TOSOH CORPORATION (MANUFACTURER)
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, TOKYO 1058623
Manufacturer CountryJA
Manufacturer Postal Code1058623
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOSOH HLC-723G8 ANALYZER G8
Generic NameG8
Product CodeLPC
Date Received2018-03-12
Model NumberG8
Catalog Number021560
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age5 YR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTOSOH CORPORATION
Manufacturer AddressSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, TOKYO 1058623 JA 1058623


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.