AVL CALCIUM ELECTRODE 03110354180

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-13 for AVL CALCIUM ELECTRODE 03110354180 manufactured by Roche Diagnostics.

Event Text Entries

[102750606] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[102750607] The customer received erroneous avl ca++ electrode (ca++) results for 3 patient samples on an electrolyte analyzer w/o starterkit 9180 (avl) serial number (b)(4). For patient 1 the initial ca++ result was 12. 0 mg/dl and the repeat result was 12. 35 mg/dl. The ca++ result was 1. 18 mg/dl from the same sample on an abl800 gasometer. For patient 2 on (b)(6) 2018 the initial ca++ result of 12. 89 mg/dl. The ca++ result was 1. 06 mg/dl from the same sample on an abl800 gasometer. A new sample was collected and the ca++ result of 11. 5 mg/dl. The ca++ result was 1. 02 mg/dl with the new sample on an abl800 gasometer. For patient 3 on (b)(6) 2018 the initial ca++ result of 15. 18 mg/dl. The ca++ result was 1. 43 mg/dl from the same sample on an abl800 gasometer. The results on the abl800 gasometer were considered to be correct. No erroneous results were reported outside of the laboratory. There was no allegation of an adverse event. The qc from both methods were within the expected range. There were no discrepancies observed for other patient samples.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-00786
MDR Report Key7333427
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-13
Date of Report2018-05-01
Date of Event2018-02-15
Date Mfgr Received2018-02-23
Date Added to Maude2018-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1AVL SCIENTIFIC CORP.
Manufacturer Street33 MANSELL COURT NA
Manufacturer CityROSWELL GA 30076
Manufacturer CountryUS
Manufacturer Postal Code30076
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameAVL CALCIUM ELECTRODE
Generic NameELECTRODE, ION SPECIFIC, CALCIUM
Product CodeJFP
Date Received2018-03-13
Model NumberNA
Catalog Number03110354180
Lot Number21573547
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.