CHROMID? C. DIFFICILE AGAR 43871

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-13 for CHROMID? C. DIFFICILE AGAR 43871 manufactured by Biomerieux Spain.

Event Text Entries

[102350199] A customer in (b)(6) reported a result of no growth, for a clostridium difficile quality control (qc) strain (b)(4)), in association with chromid? C. Difficile agar. The customer stated they inoculated a new lot of chromid? C. Difficile agar (lot 1006170510) with the qc strain, and had observed no bacterial growth after incubation at 36? C (+/- 1? C) in anaerobic atmosphere (generated by anaerobic jars). The customer stated that they perform the qc test for each new lot, and that clinical samples were not inoculated on this lot. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00036
MDR Report Key7334715
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-13
Date of Report2018-05-18
Date Mfgr Received2018-04-26
Date Added to Maude2018-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SPAIN
Manufacturer StreetISAAC NEWTON, 6
Manufacturer CityTRES CANTOS, MADRID 28760
Manufacturer CountrySP
Manufacturer Postal Code28760
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? C. DIFFICILE AGAR
Generic NameCHROMID? C. DIFFICILE AGAR
Product CodeJSI
Date Received2018-03-13
Catalog Number43871
Lot Number1006170510
Device Expiration Date2018-04-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SPAIN
Manufacturer AddressISAAC NEWTON, 6 TRES CANTOS, MADRID 28760 SP 28760


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.