AMSORB PLUS CAN-CAN ABSORBER * AMAB3800

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-06-30 for AMSORB PLUS CAN-CAN ABSORBER * AMAB3800 manufactured by Armstrong Medical Limited.

Event Text Entries

[466031] The amsorb can-can absorber is marketed as a substitue for the ge disposable multi absorber, ref 8003138. We have been using them with the ge aisys anesthesia machine. In several instances, we have encountered extreme difficulty mounting the amsorb unit on the aisys machine. It would appear that inside diameter of the water trap portion (and possibly the other ports) are too small, making it difficult to attach and possibly damaging the o-ring. Comparing of the amsorb canister to the ge canister, the o-ring that seals the water trap is visibly much more compressed by the amsorb canister. In addition, we have had problems with the canister becoming dislodged during use. The lugs that hold the canister in place appear to be shorter than the ge version, contributing to the problem. In several cases, the canister "popped off" during the leak test. The diameter of the amsorb canister is also much larger, and touches against part of the anesthesia machine, possibly preventing proper alignment of the ports and locking mechanism. The device is suspected of causing circuit leaks in some cases. It is also possible to mount the canister backward in the machine and this has been observed in one case.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number733480
MDR Report Key733480
Date Received2006-06-30
Date of Report2006-06-30
Date of Event2006-06-29
Report Date2006-06-30
Date Reported to FDA2006-06-30
Date Added to Maude2006-07-13
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameAMSORB PLUS CAN-CAN ABSORBER
Generic NameANESTHESIA ABSORBER
Product CodeCBL
Date Received2006-06-30
Model Number*
Catalog NumberAMAB3800
Lot Number100406F31
ID Number*
OperatorPHYSICIAN
Device AvailabilityY
Device Age1 DY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key721600
ManufacturerARMSTRONG MEDICAL LIMITED
Manufacturer AddressREGULATORY INSIGHT 13 RED FOX LANE LITTLETON CO 80127 US

Device Sequence Number: 2

Brand NameAISYS ADU CARESTATION
Generic NameANESTHESIA MACHINE
Product CodeBSZ
Date Received2006-06-30
Model Number1011-9000-000
Catalog Number1011-9000-000
Lot NumberNONE
ID Number*
Device AvailabilityY
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key721601
ManufacturerDATEX-OHMEDA, INC
Manufacturer Address3030 OHMEDA DRIVE MADISON WI 537077550 US


Patients

Patient NumberTreatmentOutcomeDate
10 2006-06-30

© 2026 FDA.report