C8313, XXS ALEXIS WND PROT/RET 5/BX 101355501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other,use report with the FDA on 2018-03-13 for C8313, XXS ALEXIS WND PROT/RET 5/BX 101355501 manufactured by Applied Medical Resources.

Event Text Entries

[102586520] No product is being returned for evaluation, but a lot # is provided. A device history report is to be reviewed by engineering. A final report will be sent once the results have been analyzed.
Patient Sequence No: 1, Text Type: N, H10


[102586521] Procedure performed: unk. (b)(4) received via maude database: (b)(4). Event description: "when removing a portion of the stomach via the port site, the alexis wound protector broke. A portion of the wound protector was in the abdomen. The surgeon removed the complete broken portion that was in the abdomen. X-ray was obtained. X-ray was negative. " type of intervention: "the surgeon removed the complete broken portion that was in the abdomen. " patient status: unk.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027111-2018-00103
MDR Report Key7335029
Report SourceHEALTH PROFESSIONAL,OTHER,USE
Date Received2018-03-13
Date of Report2018-03-19
Date of Event2018-01-10
Date Mfgr Received2018-02-12
Device Manufacturer Date2017-09-01
Date Added to Maude2018-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal92688
Manufacturer Phone9497138059
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameC8313, XXS ALEXIS WND PROT/RET 5/BX
Generic NameKGW
Product CodeKGW
Date Received2018-03-13
Model NumberC8313
Catalog Number101355501
Lot Number1306143
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAPPLIED MEDICAL RESOURCES
Manufacturer Address22872 AVENIDA EMPRESA RANCHO SANTA MARGARITA CA 92688 US 92688


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.