ORS-100N WARMER DRAPE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-03-13 for ORS-100N WARMER DRAPE manufactured by Microtek Dominicana, S.a..

Event Text Entries

[102327545] The sample received was evaluated. A small cut/hole was observed about 5 1/2 inches from the center of the drape. This looks to have been caused by a sharp of some sort. The dhr was reviewed and it was noticed that this lot had 95 cases that were manufactured from 06/21/2016 to 06/22/2016. No defects were reported during quality inspections. This is a non sterile part that is supplied to a kit packer for further packaging and sterilization. Based on the device history record and sample review, this does not appear to be the result of a personnel, process or material issue. Because this appears to be related to misuse by the customer, no actions are being taken at this time.
Patient Sequence No: 1, Text Type: N, H10


[102327546] A customer is reporting a complaint from a hospital stating that they received a fluid warmer drape that contained a hole within the drape. The sample was received by the customer and the reported concern was confirmed. A small tear within the plastic was noted and upon examining the location of the drape, it was noted that there was nothing sharp next to the drape that could have caused such a tear. Also, the drape is received folded and is placed within the kit as such.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043817-2018-00007
MDR Report Key7335774
Report SourceDISTRIBUTOR
Date Received2018-03-13
Date of Report2018-02-19
Date Mfgr Received2018-02-19
Date Added to Maude2018-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK DOMINICANA
Manufacturer StreetZONA FRANCA NO. 2
Manufacturer CityLA ROMANA,
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORS-100N WARMER DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2018-03-13
Returned To Mfg2018-03-05
Model NumberORS-100N
Lot NumberD161731
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK DOMINICANA, S.A.
Manufacturer AddressZONA FRANCA NO. 2 LA ROMANA, DR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-13