ENDOVIVE? JEJUNAL FEEDING TUBE M00566340 6634

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2018-03-13 for ENDOVIVE? JEJUNAL FEEDING TUBE M00566340 6634 manufactured by Boston Scientific - Costa Rica (coyol).

Event Text Entries

[102320141] The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown. However, the complainant reported that the device was not expired. (b)(4). The device has been received for analysis. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[102320142] It was reported to boston scientific corporation that an endovive three-port through the peg jejunal tube was used during a percutaneous endoscopic gastrostomy procedure performed on (b)(6) 2018. According to the complainant, after placement was performed, tube blockage occurred frequently due to kink, and the kink issue was resolved using a biopsy forceps through endoscope. Recurrence of issue within a week was noticed. The procedure was completed with a new endovive three-port through the peg jejunal tube. There were no patient complications reported due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-00767
MDR Report Key7335969
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2018-03-13
Date of Report2018-02-26
Date of Event2018-02-23
Date Mfgr Received2018-03-25
Date Added to Maude2018-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Manufacturer Street2546 FIRST STREET PROPARK FREE ZONE
Manufacturer CityALAJUELA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDOVIVE? JEJUNAL FEEDING TUBE
Generic NameTUBE, FEEDING
Product CodeFPD
Date Received2018-03-13
Returned To Mfg2018-03-05
Model NumberM00566340
Catalog Number6634
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - COSTA RICA (COYOL)
Manufacturer Address2546 FIRST STREET PROPARK FREE ZONE ALAJUELA CS


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.