VARADY PHLEBECTOMY EXTRACTOR, 6-3/4 SM HOOK 10310

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2018-03-14 for VARADY PHLEBECTOMY EXTRACTOR, 6-3/4 SM HOOK 10310 manufactured by Hebumedical Gmbh.

Event Text Entries

[102591652] The device was returned without the broken-off tip. It had been manufactured in november 2014. Review of manufacturing records confirmed device conformance to specs at time of release. Hardness of returned device was tested and found conforming to specifications. The device was subject to microscopic examination. It showed no evidence of extensive reuse. Since the device was returned incomplete, i. E. , without the broken-off tip, it is not possible to reach a definite conclusion with regard to the cause of the fracture. However, this type of extractor has a very delicate hook tip due to the natural anatomical requirements of the vein-stripping procedure, and the tip may fracture if excessive pressure is applied by the surgeon.
Patient Sequence No: 1, Text Type: N, H10


[102591653] During an unspecified procedure, the hook of the varady phlebectomy extractor broke off and was stuck inside the extracted vein. No patient injury nor delay occurred nor were any revisions or medical interventions required.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611112-2018-00002
MDR Report Key7337847
Report SourceOTHER
Date Received2018-03-14
Date of Report2018-03-14
Date of Event2017-12-21
Date Mfgr Received2018-02-15
Device Manufacturer Date2014-11-05
Date Added to Maude2018-03-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. THOMAS BUTSCH
Manufacturer StreetBADSTRASSE 8
Manufacturer CityTUTTLINGEN, 78532
Manufacturer CountryGM
Manufacturer Postal78532
Manufacturer G1HEBUMEDICAL GMBH
Manufacturer StreetBADSTRASSE 8
Manufacturer CityTUTTLINGEN, 78532
Manufacturer CountryGM
Manufacturer Postal Code78532
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVARADY PHLEBECTOMY EXTRACTOR, 6-3/4 SM HOOK
Generic NameEXTRACTOR
Product CodeGAI
Date Received2018-03-14
Returned To Mfg2018-02-22
Catalog Number10310
Lot Number100088-1411
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHEBUMEDICAL GMBH
Manufacturer AddressBADSTRASSE 8 TUTTLINGEN, 78532 GM 78532


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.