TC TENOTOMY SCISSDELCVD1/SERR130MM DO250R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-03-15 for TC TENOTOMY SCISSDELCVD1/SERR130MM DO250R manufactured by Aesculap Ag.

Event Text Entries

[102582680] (b)(4). Manufacturing site evaluation: the tip of the scissors are broken. Pictorial documentation was carried out with a digital microscope. Hardness measurement was carried out with hardness tester. The device history files have been checked and were found to be according to the specifications. Based on the information available as well as a result of the investigation the root cause of the failure is most probably related to an insufficient usage. This kind of instrument is designated for delicate use only. It is liable that a mechanical overload situation led to the breakage. The visual investigation and the hardness test indicated that the quality requirements are in the specified tolerance range. A capa is not necessary.
Patient Sequence No: 1, Text Type: N, H10


[102582681] Country of complaint: usa. Customer states that during intra-operative surgery the tip of scissors broke off. No surgical delay and no patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610612-2018-00100
MDR Report Key7343488
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-03-15
Date of Report2018-03-15
Date of Event2016-08-29
Date Facility Aware2018-02-27
Date Mfgr Received2016-08-31
Device Manufacturer Date2016-04-19
Date Added to Maude2018-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTC TENOTOMY SCISSDELCVD1/SERR130MM
Generic NameDENTAL SCISSORS
Product CodeEGN
Date Received2018-03-15
Returned To Mfg2016-11-03
Model NumberDO250R
Catalog NumberDO250R
Lot Number4507434213
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age23 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-03-15

© 2024 FDA.report