API? 20E TEST STRIP 20100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-16 for API? 20E TEST STRIP 20100 manufactured by Biomerieux Sa.

Event Text Entries

[103157940] A customer in (b)(6) contacted biom? Rieux to report the occurrence of a misidentification of raoultella planticola (external qc strain) as klebsiella pneumonia in association with the api? 20e test. No alternate test method or confirmatory testing was performed by the customer. The external qc testing was performed one year ago; the strain is no longer available for retest. Prior to 2001, the name of raoultella planticola was klebsiella planticola. The biochemical characteristics between raoultella planticola and klebsiella pneumonia are similar. It is important to note that raoultella planticola is not in the api? 20e database; therefore, it is not possible to identify raoultella planticola theapi? 20e test strip. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There was no patient directly associated with the qc strain. Biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00042
MDR Report Key7344955
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-16
Date of Report2018-04-23
Date Mfgr Received2018-03-27
Date Added to Maude2018-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELLEN WELTMER
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SA
Manufacturer Street3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE 383
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPI? 20E TEST STRIP
Generic NameAPI? 20E TEST STRIP
Product CodeJSS
Date Received2018-03-16
Catalog Number20100
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE 383 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-16