10-20 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-03-20 for 10-20 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-50 manufactured by Charter Medical, Ltd..

Event Text Entries

[103281142] A single, used sample was received for investigation. The bag was charged with compressed air and leak tested under water; a leak was discovered from the right top corner of the bag. The area where the observed leak originated was examined, using a magnification inspection lamp, and a hole was verified in the film. The examination also revealed a film off-set issue in the perimeter weld at the port section of the bag. The offset is indicative of a film mis-alignment issue from the manufacturing process. During the manufacturing process, the bag is placed into tooling for a welding process, which requires manual film alignment. As a result of the film misalignment observed on the returned sample the manufacturing tooling likely placed stress on the film at the hole location and produced a tear, or weak area, in the film.
Patient Sequence No: 1, Text Type: N, H10


[103281143] While performing a final product fill in a cf-50 bag, the user reported a spill was coming from a hole in the freezing bag. The product was discarded; however the bag was saved for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00006
MDR Report Key7355969
Date Received2018-03-20
Date of Report2018-03-20
Date of Event2018-02-19
Date Mfgr Received2018-02-20
Device Manufacturer Date2017-11-21
Date Added to Maude2018-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name10-20 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name10-20 ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2018-03-20
Returned To Mfg2018-03-15
Model NumberCF-50
Catalog NumberCF-50
Lot Number151497
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.