EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-03-20 for EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F manufactured by Olympus Medical Systems Corp..

Event Text Entries

[103410114] The scope was returned to olympus for evaluation. A visual inspection was performed on the returned scope and found the balloon tip on the probe unit broke off and missing. The scope passed the leak test. In addition, the insertion tube was noted to be buckled at the distal end area near the bending cover glue. In addition, the biopsy channel port has deep scratches. The scope was repaired and returned to the user facility. The cause of the reported event could not be conclusively determined as limited information was provided by the user facility. However, based on similar related events, the most probable cause of the reported event is improper handling due to excessive force applied when attaching the balloon.
Patient Sequence No: 1, Text Type: N, H10


[103410115] Olympus was informed that during a bronchoscopy procedure, a piece of the scope distal end broke off and fell inside the patient? S right lung. The broken piece was retrieved. No patient harm was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2018-00198
MDR Report Key7356325
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-03-20
Date of Report2018-03-20
Date of Event2018-02-27
Date Mfgr Received2018-02-27
Date Added to Maude2018-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone408935-512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Generic NameULTRASONIC BRONCHOFIBERVIDEOSCOPE
Product CodePSV
Date Received2018-03-20
Returned To Mfg2018-03-01
Model NumberBF-UC180F
Catalog NumberBF-UC180F
Lot NumberN/A
ID Number04953170356360
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.