KODAK DIRECTVIEW CR CASSETTE 890-7321

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2006-07-01 for KODAK DIRECTVIEW CR CASSETTE 890-7321 manufactured by Eastman Kodak Company.

Event Text Entries

[472498] A radiograph taken for a trauma case showed a foreign body. Additional radiographs were taken to locate the foreign body and these showed no foreign body.
Patient Sequence No: 1, Text Type: D, B5


[7779128] Kodak received the screen and cassette for evaluation. Our evaluation included creating and printing flat field images and a physical examination of the screen. The flat field images were printed at an extreme contrast in order to ehance any image artifacts. The flat field image showed artifacts that were classified as fingerprint artifacts. These were located at the edges and corners of the image. The physical examination showed an area with a small abrasion at the bottom of the screen. In the flat field image this looked like a foreign body. The fingerprint artifacts have been seen on other cr images and are associated with handling of the screens by bare hands. Many lotions and waterless hand cleaners contain chemicals that can damage phosphor screens. The affected cr screen was removed and replaced with a new cr screen.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1317307-2006-00006
MDR Report Key736446
Report Source05,06
Date Received2006-07-01
Date of Report2006-06-29
Date of Event2006-01-18
Date Mfgr Received2006-06-08
Device Manufacturer Date2004-09-01
Date Added to Maude2006-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactJOHN PARDO
Manufacturer Street343 STATE STREET
Manufacturer CityROCHESTER NY 146501131
Manufacturer CountryUS
Manufacturer Postal146501131
Manufacturer Phone5857245974
Manufacturer G1EASTMAN KODAK COMPANY
Manufacturer Street1041 RIDGE ROAD WEST BUILDING 214
Manufacturer CityROCHESTER NY 14652349
Manufacturer CountryUS
Manufacturer Postal Code14652 3491
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKODAK DIRECTVIEW CR CASSETTE
Generic NameCR RIGID CASSETTE
Product CodeIXA
Date Received2006-07-01
Returned To Mfg2006-02-23
Model NumberNA
Catalog Number890-7321
Lot Number1814
ID Number1814005252CR9834
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key724335
ManufacturerEASTMAN KODAK COMPANY
Manufacturer Address343 STATE STREET ROCHESTER NY 14650 US
Baseline Brand NameKODAK DIRECTVIEW CR CASSETTE
Baseline Generic NameCR CASSETTE WITH ATTACHED SP SCREEN
Baseline Model NoNA
Baseline Catalog No890-7321
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-07-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.