BIRD HIGH FLOW BLENDER SERIES 03800A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-03-26 for BIRD HIGH FLOW BLENDER SERIES 03800A manufactured by Vyaire Medical, Inc.

Event Text Entries

[103460857] At this time, vyaire has not received the suspect device from the customer for evaluation. Several attempts have been made to contact the customer for additional information and to evaluate the suspect device for the reported issue. At this time, the customer has not responded to any contact attempts and vyaire medical is unable to confirm the customer's reported device issue. In the event the device is received for evaluation or additional information is received a follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[103460858] The customer reported the bird high flow blender series was set to dispense oxygen at fifty-five percent. The customer reported the oxygen percentage was analyzed with a calibrated analyzer and was shown to be seventy-seven percent. The customer reported the blender was removed from service and was replaced with another blender. The issue occurred during patient-use; the customer reported there is no patient consequence associated with the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2018-07611
MDR Report Key7370489
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-03-26
Date of Report2018-03-26
Date of Event2017-11-01
Date Mfgr Received2018-03-05
Device Manufacturer Date2011-01-01
Date Added to Maude2018-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN GRAF
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer G1VYAIRE MEDICAL, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIRD HIGH FLOW BLENDER SERIES
Generic NameMIXER, BREATHING GASES, ANESTHESIA INHALATION
Product CodeBZR
Date Received2018-03-26
Model NumberBLENDER
Catalog Number03800A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-03-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.