NEXTGEN PLASMA MANAGEMENT SYSTEM NEXTGEN 4.0.0P2 - 4.1.0

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2018-03-26 for NEXTGEN PLASMA MANAGEMENT SYSTEM NEXTGEN 4.0.0P2 - 4.1.0 manufactured by Haemonetics Corporation Software Solutions.

Event Text Entries

[103682768] Haemonetics has identified an issue with the nextgen software, when using the swap-collection or change-collection-donor features of nextgen. These features allow customers to correct their records when staff make mistakes in recording collections. The error occurs if one of the donors involved in the swap or change-collection operation has had an rbc loss. Due to the error, a rbc deferral would not be created for the correct donor. Haemonetics has released a technical bulletin to the haemonetics download center on march 16 2018, which notifies haemonetics field services of the potential software issue which impacts only two customer sites. Additionally, haemonetics is going to release an updated version of the nextgen software which will correct this issue in the software. The patch is planned to be released during the summer of 2018.
Patient Sequence No: 1, Text Type: N, H10


[103682769] On (b)(4) 2018, internal testing found an error when using the swap-collection or change-collection-donor features of nextgen. These features allow customers to correct their records when staff make mistakes in recording collections. The error occurs if one of the donors involved in the swap or change-collection operation has had an rbc loss. Due to the error, a rbc deferral would not be created for the correct donor. This issue was reviewed on feb 28, 2018 and determined to be a potential health and safety risk. If a donor has an rbc loss that should result in a deferral, and the collection was involved in a swap or change, the deferral may be applied to the wrong donor.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951268-2018-00003
MDR Report Key7370881
Report SourceOTHER
Date Received2018-03-26
Date of Report2018-03-26
Date of Event2018-02-27
Date Mfgr Received2018-02-27
Date Added to Maude2018-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR DAVID RAMSAY
Manufacturer Street400 WOOD ROAD
Manufacturer CityBRAINTREE MA 02184
Manufacturer CountryUS
Manufacturer Postal02184
Manufacturer Phone7813487327
Manufacturer G1HAEMONETICS CORPORATION SOFTWARE SOLUTIONS
Manufacturer Street4925 ROBERT J. MATTHEWS PARKWA SUITE 100
Manufacturer CityEL DORADO HILLS CA 95762
Manufacturer CountryUS
Manufacturer Postal Code95762
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEXTGEN PLASMA MANAGEMENT SYSTEM
Generic NameD ESTABLISHMENT COMPUTER SOFTWARE AND ACCESSORIES
Product CodeMMH
Date Received2018-03-26
Model NumberNEXTGEN 4.0.0P2 - 4.1.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHAEMONETICS CORPORATION SOFTWARE SOLUTIONS
Manufacturer Address4925 ROBERT J. MATTHEWS PARKWA SUITE 100 EL DORADO HILLS CA 95762 US 95762


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-26

© 2024 FDA.report