SECURA NO-STING BARRIER FILM 3MLX50SWABS 66800711

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-03-28 for SECURA NO-STING BARRIER FILM 3MLX50SWABS 66800711 manufactured by Smith & Nephew Medical Ltd..

Event Text Entries

[103643241]
Patient Sequence No: 1, Text Type: N, H10


[103643242] It was reported that secure was used on the wound edge when using npwt system. Small wound leakages noticed and electrode switchpen used to stop bleeding. Vapor was lit on fire. No patient harm reported.
Patient Sequence No: 1, Text Type: D, B5


[107863077] The device was not evaluated due to samples were not returned for investigation. Investigation results: we have now concluded our investigation into this report of skin prep no sting wipes. It can be confirmed that each batch released to market undergoes full testing as outlined in the finished product specification, and must comply with all the requirements. The associated skin prep no sting wipe product information for use (ifu) leaflet states the following: secura no-sting barrier films is flammable: use in a well-ventilated area. Avoid using around flames and sources of ignition. Keep out of the reach of children. For external use only. Additionally, complaints of a similar nature will be closely monitored for any adverse trends and further action considered.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8043484-2018-00072
MDR Report Key7376527
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-03-28
Date of Report2018-05-09
Date of Event2018-03-02
Date Mfgr Received2018-03-05
Date Added to Maude2018-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR MARKUS POETTKER
Manufacturer StreetSCHACHENALLEE 29
Manufacturer CityAARAU 5001
Manufacturer CountrySZ
Manufacturer Postal5001
Manufacturer G1SMITH & NEPHEW MEDICAL LTD.
Manufacturer Street101 HESSLE ROAD
Manufacturer CityHULL HU3 28N
Manufacturer CountryUK
Manufacturer Postal CodeHU3 28N
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSECURA NO-STING BARRIER FILM 3MLX50SWABS
Generic NameTAPE AND BANDAGE, ADHESIVE
Product CodeKMF
Date Received2018-03-28
Catalog Number66800711
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW MEDICAL LTD.
Manufacturer Address101 HESSLE ROAD HULL HU3 28N UK HU3 28N


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-28

© 2026 FDA.report