HUDSON TUBING, OXYGEN SUPPLY 7' 1115

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2018-03-28 for HUDSON TUBING, OXYGEN SUPPLY 7' 1115 manufactured by Teleflex Medical.

Event Text Entries

[103743663] (b)(4). Visual inspection of the product involved in the complaint was performed on two pictures provided by the customer. Some marks on the tube were observed, however, no issues can be observed that can lead this customer complaint. In order to perform a proper and thorough investigation to confirm the alleged defect reported and determine a root cause, it is necessary to evaluate the sample involved in this complaint. Device sample has not been returned at the time of this report. The device history record (dhr) has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint. The dhr shows that the product was assembled and inspected according to our specifications. Customer complaint cannot be confirmed based only on the information provided. If the sample becomes available this investigation will be updated with the evaluation results. Teleflex will continue to monitor and trend for complaints of this nature.
Patient Sequence No: 1, Text Type: N, H10


[103743664] Customer complaint alleges "tubing kinks where it comes off the wall and collapses making it useless to deliver o2 to patient. " alleged defect reported as detected during use. There was no report of patient impact or consequence. Customer submitted photos with the complaint.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2018-00086
MDR Report Key7379060
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2018-03-28
Date of Report2018-03-14
Date of Event2018-03-13
Date Mfgr Received2018-05-18
Device Manufacturer Date2017-11-01
Date Added to Maude2018-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON TUBING, OXYGEN SUPPLY 7'
Generic NameTUBING, PRESSURE AND ACCESSORI
Product CodeBYX
Date Received2018-03-28
Returned To Mfg2018-05-07
Catalog Number1115
Lot Number74L1700135
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-28

© 2024 FDA.report