TOTAL PROTEIN URINE/CSF GEN.3 03333825190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-03-30 for TOTAL PROTEIN URINE/CSF GEN.3 03333825190 manufactured by Roche Diagnostics.

Event Text Entries

[104258735] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[104258736] The customer complained of questionable tpuc3 total protein urine/csf gen. 3 for multiple patient urine samples on a cobas 6000 c (501) module serial number (b)(4) when compared to a hitachi 7180 and the pyrogallol red method. After the customer began using the c 501 with the tpuc3 reagent a physician questioned the increase in the urinary protein and creatinine ratio. Of the data provided, there were 69 urine samples with erroneous tpuc results. Refer to the attachment to the medwatch for all patient data. Erroneous results were reported outside of the laboratory. There was no allegation of an adverse event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-00977
MDR Report Key7385744
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-03-30
Date of Report2018-05-04
Date of Event2018-03-13
Date Mfgr Received2018-03-13
Date Added to Maude2018-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTOTAL PROTEIN URINE/CSF GEN.3
Generic NameTURBIDIMETRIC, TOTAL PROTEIN
Product CodeJGQ
Date Received2018-03-30
Model NumberNA
Catalog Number03333825190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.