PR4100 MICROPLATE READER US 45591587

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-03-30 for PR4100 MICROPLATE READER US 45591587 manufactured by Bio-rad Laboratories.

Event Text Entries

[104277510]
Patient Sequence No: 1, Text Type: N, H10


[104277531] On (b)(6) 2018 bio-rad laboratories technical support received a call from a customer stating that the d-link dub-h4 usb hub, used with the pr4100 microplate reader system, started smoking when the user turned on the pr4100 computer. The pr4100 microplate reader is a 96-well absorbance reader for the measurement of light absorbance (optical density-od) of liquid media. The instrument is intended to be used primarily for in vitro diagnostic analysis of samples from the human body to obtain information on physiological and pathological states. The pr4100 system was turned off and no injury occurred. Bio-rad laboratories technical support sent the customer a tripp lite usb hub as a replacement. A field corrective action has been generated to replace all existing d-link hubs that are currently in use for pr4100 microplate reader customers. Bio-rad laboratories will submit a follow-up report upon completion of the field corrective action.
Patient Sequence No: 1, Text Type: D, B5


[131389927] As noted in our initial report bio-rad laboratories committed to providing a follow-up report to the fda. There has been no new or additional information, and bio-rad laboratories has completed the field action for this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3022521-2018-00001
MDR Report Key7386745
Date Received2018-03-30
Date of Report2018-12-21
Date of Event2018-02-17
Date Mfgr Received2018-03-01
Date Added to Maude2018-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DAVID BHEND
Manufacturer Street6565 185TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4254981758
Manufacturer G1BIO-RAD LABORATORIES
Manufacturer Street3, BOULEVARD RAYMOND POINCARE
Manufacturer CityMARNES-LA-COQUETTE, 92430
Manufacturer CountryFR
Manufacturer Postal Code92430
Single Use3
Previous Use Code3
Removal Correction Number3022521-4/2/2018-002C
Event Type3
Type of Report0

Device Details

Brand NamePR4100 MICROPLATE READER US
Generic NameSPECTROPHOTOMETER FOR CLINICAL USE
Product CodeJJQ
Date Received2018-03-30
Returned To Mfg2018-02-27
Catalog Number45591587
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIO-RAD LABORATORIES
Manufacturer Address3, BOULEVARD RAYMOND POINCARE MARNES-LA-COQUETTE, 92430 FR 92430


Patients

Patient NumberTreatmentOutcomeDate
10 2018-03-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.