ELECSYS CK-MB STAT IMMUNOASSAY 05957648190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-04-03 for ELECSYS CK-MB STAT IMMUNOASSAY 05957648190 manufactured by Roche Diagnostics.

Event Text Entries

[104426588] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[104426589] The customer received questionable elecsys ck-mb stat immunoassay results when comparing reagent lots. Of the data provided, only the results for one patient sample were discrepant. The result with reagent lot 222122 was 58 ng/ml and the result with reagent lot 297009 was 49. 98 ng/ml. The results with reagent lot 222122 had been released outside of the laboratory to medical personnel. When the samples were repeated with the new reagent lot, those results were seen as erroneous and were not released. There was no allegation of an adverse event. The type and serial number of analyzer used for the testing was requested but was not provided. The investigation was unable to find a definitive root cause. An issue with the single reagent pack was suspected by the customer as there were microparticles in the lid most likely due to transportation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01005
MDR Report Key7391506
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-04-03
Date of Report2018-04-03
Date of Event2018-03-14
Date Mfgr Received2018-03-15
Date Added to Maude2018-04-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS CK-MB STAT IMMUNOASSAY
Generic NameCOLORIMETRIC METHOD, CPK OR ISOENZYMES
Product CodeJHY
Date Received2018-04-03
Model NumberNA
Catalog Number05957648190
Lot Number297009
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-03