CYSTOSCOPY IRRIGATION SET 2C4040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2018-04-03 for CYSTOSCOPY IRRIGATION SET 2C4040 manufactured by Baxter Healthcare - Cartago.

Event Text Entries

[104443520] Contact office: (b)(4). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[104443521] It was reported that a hair was discovered within the packaging of a cysto/bladder irrigation set. This event was discovered during an unspecified process step before patient use. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[117736001] The device was received for evaluation. A batch review was conducted and there were no deviations found related to the reported condition during the manufacture of this lot. Visual inspection was performed on the returned unit via the naked eye. The reported condition was verified due to a loose particle found inside the product pouch. The particle was concluded to be a hair. The possible cause was incorrect manufacturing practices during the manual assembly process. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2018-01836
MDR Report Key7394190
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2018-04-03
Date of Report2018-05-08
Date of Event2018-03-08
Date Mfgr Received2018-04-30
Date Added to Maude2018-04-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CARTAGO
Manufacturer StreetSEE H10 SEE H10
Manufacturer CityCARTAGO 30106
Manufacturer CountryCS
Manufacturer Postal Code30106
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCYSTOSCOPY IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2018-04-03
Returned To Mfg2018-04-09
Model NumberNA
Catalog Number2C4040
Lot NumberNI
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CARTAGO
Manufacturer AddressCARTAGO


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-03

© 2024 FDA.report