CLINIMACS TUBING SET TS 73105 161-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-04 for CLINIMACS TUBING SET TS 73105 161-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[104275074]
Patient Sequence No: 1, Text Type: N, H10


[104275075] During the integrity test run just after the ts kit was loaded onto the clinimacs machine, a small leak from one of the junctions under the liquid sensor was noted. The run was terminated upon the failed kit inspection. The lot number of the ts kit was b2398. A new ts kit of the same lot (b2398) was loaded onto the clinimacs machine, passed inspection and the selection was performed without incident. Product was not affected negatively in any way and processing continued according to relevant standard operating procedures. Miltenyi was notified via email of the kit leak.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7395234
MDR Report Key7395234
Date Received2018-04-04
Date of Report2018-03-20
Date of Event2018-03-06
Report Date2018-03-20
Date Reported to FDA2018-03-20
Date Reported to Mfgr2018-03-20
Date Added to Maude2018-04-04
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINIMACS TUBING SET TS
Generic NameMAGNETIC CELL SELECTION SYSTEM FOR TREATMENT OF ACUTE MYELOGENOUS LEUKEMIA (AML
Product CodeOVG
Date Received2018-04-04
Model Number73105
Catalog Number161-01
Lot NumberB2398
Device Expiration Date2018-05-23
Device AvailabilityN
Device Age7 MO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer Address2303 LINDBERGH ST. AUBURN CA 95602 US 95602


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-04

© 2026 FDA.report