DSU-H 70-0285

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-04-04 for DSU-H 70-0285 manufactured by Nephros Inc.

Event Text Entries

[104542345] When initially evaluation without the product return, it was concluded that the event was not mdr reportable and was likely that nephros disinfection and installation procedures were not followed. Nonetheless, after a complete investigation there was indeed fiber breaks in both bundles of the filter which could allow bacteria to pass through. Therefore, from the information gathered on 03/05/2018, this is an reportable event.
Patient Sequence No: 1, Text Type: N, H10


[104542346] On (b)(6) 2018 it was reported that a healthcare facility received a positive culture downstream of a dsu-h (70-0285) in one of their ice machines. There was no reported adverse events or health issues connected with the positive result. Nephros dsu-h is an in-line water filter, class ii medical device that is placed before ice machines in health care facilities in order to retain bacteria, virus and endotoxin. The user facility routinely test their water at their ice machine and found that there was positive bacteria detected. From the initial discussion, it is likely that our disinfection and installation procedures were not followed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003337893-2018-00001
MDR Report Key7397667
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-04
Date of Report2018-04-04
Date of Event2018-01-25
Date Mfgr Received2018-01-25
Date Added to Maude2018-04-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. VASHONE THOMAS
Manufacturer Street380 LACKAWANNA PLACE
Manufacturer CitySOUTH ORANGE NJ 07079
Manufacturer CountryUS
Manufacturer Postal07079
Manufacturer Phone2013450829
Manufacturer G1NEPHROS, INC.
Manufacturer Street380 LACKAWANNA
Manufacturer CitySOUTH ORANGE NJ 07079
Manufacturer CountryUS
Manufacturer Postal Code07079
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDSU-H
Generic NameSYSTEM, WATER PURIFICATION, GENERAL MEDICAL U, PRODUCT CODE: NHV
Product CodeNHV
Date Received2018-04-04
Returned To Mfg2018-02-07
Model Number70-0285
Lot NumberPI17/0342
OperatorBIOMEDICAL ENGINEER
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNEPHROS INC
Manufacturer Address380 LACKAWANNA PLACE SOUTH ORANGE NJ 07079 US 07079


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.