NEOBLUE COZY 001354

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-04-04 for NEOBLUE COZY 001354 manufactured by Natus Medical Inc..

Event Text Entries

[104544345] The tsr inquired how the measurement was taken, and the customer confirmed the meter sensor was placed over the bubble mattress without the disposable cover. The tsr then suggested to the customer to add the cover and to retake measurements, per manual instructions. The tsr also instructed the customer on how to adjust the intensity using the potentiometers on the unit. On 7 mar 2018, the tsr followed up with the customer on the status of the device and left a message to the customers. On 16 mar 2018, the tsr contacted the customer again to follow up, and the customer confirmed that they were successfully able to adjust the intensity to specifications.
Patient Sequence No: 1, Text Type: N, H10


[104544346] The customer reported to a natus technical service representative (tsr) that their neoblue cozy unit was reading high (43. 1 microw/cm2/nm) when measured with a natus radiometer. No death, serious injury, or harm to patients/users has been reported. No delays in treatment or environmental concerns have been reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00112
MDR Report Key7398451
Report SourceUSER FACILITY
Date Received2018-04-04
Date of Report2018-03-06
Date Mfgr Received2018-03-06
Date Added to Maude2018-04-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685133
Manufacturer G1NATUS MEDICAL INC.
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEOBLUE COZY
Generic NameNEOBLUE COZY
Product CodeLBI
Date Received2018-04-04
Model Number001354
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC.
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-04

© 2026 FDA.report