ELECSYS TOTAL PSA IMMUNOASSAY 04491734160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-04-05 for ELECSYS TOTAL PSA IMMUNOASSAY 04491734160 manufactured by Roche Diagnostics.

Event Text Entries

[104840938] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[104840939] The customer received questionable elecsys total psa immunoassay results for one patient compared to a siemens vista analyzer. Refer to the attachment to the medwatch for all patient data. The total psa was reported as <0. 1 ng/ml. The customer was not aware that the patient was treated based on the results and he did not know if the patient had his prostate removed. There was no allegation of an adverse event. The customer performed a biotin study which showed no interference. The sample had no change in results when treated with streptavidin. The customer declined a service visit as they did not believe there was an analyzer issue. He stated the cobas e601 results were probably correct as they were supported by a similar result from the beckman analyzer. The customer used cobas 6000 e 601 module serial number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01059
MDR Report Key7398697
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-05
Date of Report2018-04-25
Date of Event2018-03-01
Date Mfgr Received2018-03-13
Date Added to Maude2018-04-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS TOTAL PSA IMMUNOASSAY
Generic NameTOTAL,PROSTATE SPECIFIC ANTIGEN(NONCOMPLEXED&COMPLEXED) FOR DETECTION OF PROS
Product CodeMTF
Date Received2018-04-05
Model NumberNA
Catalog Number04491734160
Lot Number21602701
ID NumberNA
Device Expiration Date2018-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-05

© 2026 FDA.report